The Service Information provides the following information:
Precautions and safety information
• Specifications
• Basic maintenance (cleaning, replacing the battery and fuses)
• Performance test procedures
• Calibration and calibration adjustment procedures
Form Number: TM61633-5
Revision: 5. Date: December 2006
Form Number TM61633-5 Rev 5 Dec 2006
Page view 0
1 2 3 4 5 6 ... 22 23

Summary of Contents

Page 1 - MODELS: 61-633

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODELS: 61-633 61-635 The Service Information provides the following information: • Precautions and s

Page 2

Page 8 Required Equipment Required equipment is listed in Table B. If the recommended models are not available, equipment with e

Page 3

Page 9 Basic Maintenance Warning To avoid shock, remove the test leads and any input signals before opening the case or replaci

Page 4

Page 10 Form Number TM61633-5 Rev 5 Dec 2006

Page 5

Page 11 Replacing Fuses Warning To avoid electrical shock, remove the test leads and any input signals before replacing the battery o

Page 6

Page 12 Performance Tests The following performance tests verify the complete operability of the Meter and check the accuracy of each Meter function

Page 7

Page 13 Testing the Voltage Function To verify accuracy in the DC and AC voltage ranges, do the following: 1. Connect the Calibrator to the VΩ and

Page 8

Page 14 Testing Millivoltage (mV) Function To verify accuracy of the mV function, do the following. 1. Connect the Calibrator to the VΩ and COM in

Page 9

Page 15 Testing the Resistance Function To verify the accuracy of the resistance function, do the following: 1. Connect the Calibrator to VΩ and CO

Page 10

Page 16 Testing the Capacitance Function (cont’d) d) For each input, compare the readings on the Meter display to display readings in

Page 11

Page 17 Table 9. AC mA Test Input Readings Step Current Frequency 61-633 61-635 1 36.000mA 50 Hz 35.488 to 36.512 35.632 to 36.368 2 36.000mA

Page 12 - Page 10

TABLE OF CONTENTS Title PageIntroduction 1Precautions and Saf

Page 13

Page 18 Table 12. Frequency Test Input Reading Step Frequency Level 61-633 61-635 1 360Hz 0.5Vp-p 359.86 to 360.14 359.86 to 360.14 2 3.6000K

Page 14

Page 19 IDEAL 61-630 Series calibration process Calwin300 Software 1. Select communication port or 2. Select power line frequency or 3. Se

Page 15

Page 20 Figure 4 Calibration: Calibration of the meter is recommended once a year to ensure performance according to the published specifications.

Page 16

Page 21 ACV High Frequency Calibration Calibrating the ACV Frequency Response 1. 400V Range Set the calibrator to generate 200.00V / 10KHz ±0.0

Page 17

Page 1 Introduction Warning To avoid shock or injury, do not perform the verification tests or calibration procedures described in this manual unless

Page 18

Page 2 SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid p

Page 19

Page 3 SPECIFICATIONS All specifications are warranted unless noted typical and apply to models 61-633 and 61-635. Stated accuracies are at 23° ±5°C

Page 20

Page 4 Measurement Characteristics All at 23°C ±5°C, < 80% Rh. Specifications are expressed as ±([% of reading] + [number of digits]). Multiply

Page 21

Page 5 DC CURRENT: DCA 61-633 61-635 40mA, 400mA, 4A, 10A ± (0.50% + 4d) ±(0.20% + 4d) AC CURRENT: ACA 61-633 61-635 40mA, 400mA, 4A, 10A ±(1.20%

Page 22

Page 6 CAPACITANCE Capacitance 61-633 61-635 4nF, 40nF 400nF, 4µF ± (1 .90% + 20d) ± (0.90% + 20d) 40µF, 400µF ± (2.90% + 20d) ± (1 .90% + 20d) 4

Page 23

Page 7 Certifications and compliances Safety Designed to IEC 1010-1, UL3111-1 and CSA specifications 1000V DC Category II 600V DC Category III