IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODELS: 61-633 61-635 The Service Information provides the following information: • Precautions and s
Page 8 Required Equipment Required equipment is listed in Table B. If the recommended models are not available, equipment with e
Page 9 Basic Maintenance Warning To avoid shock, remove the test leads and any input signals before opening the case or replaci
Page 10 Form Number TM61633-5 Rev 5 Dec 2006
Page 11 Replacing Fuses Warning To avoid electrical shock, remove the test leads and any input signals before replacing the battery o
Page 12 Performance Tests The following performance tests verify the complete operability of the Meter and check the accuracy of each Meter function
Page 13 Testing the Voltage Function To verify accuracy in the DC and AC voltage ranges, do the following: 1. Connect the Calibrator to the VΩ and
Page 14 Testing Millivoltage (mV) Function To verify accuracy of the mV function, do the following. 1. Connect the Calibrator to the VΩ and COM in
Page 15 Testing the Resistance Function To verify the accuracy of the resistance function, do the following: 1. Connect the Calibrator to VΩ and CO
Page 16 Testing the Capacitance Function (cont’d) d) For each input, compare the readings on the Meter display to display readings in
Page 17 Table 9. AC mA Test Input Readings Step Current Frequency 61-633 61-635 1 36.000mA 50 Hz 35.488 to 36.512 35.632 to 36.368 2 36.000mA
TABLE OF CONTENTS Title PageIntroduction 1Precautions and Saf
Page 18 Table 12. Frequency Test Input Reading Step Frequency Level 61-633 61-635 1 360Hz 0.5Vp-p 359.86 to 360.14 359.86 to 360.14 2 3.6000K
Page 19 IDEAL 61-630 Series calibration process Calwin300 Software 1. Select communication port or 2. Select power line frequency or 3. Se
Page 20 Figure 4 Calibration: Calibration of the meter is recommended once a year to ensure performance according to the published specifications.
Page 21 ACV High Frequency Calibration Calibrating the ACV Frequency Response 1. 400V Range Set the calibrator to generate 200.00V / 10KHz ±0.0
Page 1 Introduction Warning To avoid shock or injury, do not perform the verification tests or calibration procedures described in this manual unless
Page 2 SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid p
Page 3 SPECIFICATIONS All specifications are warranted unless noted typical and apply to models 61-633 and 61-635. Stated accuracies are at 23° ±5°C
Page 4 Measurement Characteristics All at 23°C ±5°C, < 80% Rh. Specifications are expressed as ±([% of reading] + [number of digits]). Multiply
Page 5 DC CURRENT: DCA 61-633 61-635 40mA, 400mA, 4A, 10A ± (0.50% + 4d) ±(0.20% + 4d) AC CURRENT: ACA 61-633 61-635 40mA, 400mA, 4A, 10A ±(1.20%
Page 6 CAPACITANCE Capacitance 61-633 61-635 4nF, 40nF 400nF, 4µF ± (1 .90% + 20d) ± (0.90% + 20d) 40µF, 400µF ± (2.90% + 20d) ± (1 .90% + 20d) 4
Page 7 Certifications and compliances Safety Designed to IEC 1010-1, UL3111-1 and CSA specifications 1000V DC Category II 600V DC Category III